Sign in →

Test Code VLTS Volatile Screen, Serum

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Additional Testing Requirements


If measurement of ethylene glycol is also needed, order ETGL / Ethylene Glycol, Serum in addition to this test.



Specimen Required


Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: Full tube

Collection Instructions:

1. Do not use alcohol to clean arm. Use alternative such as Betadine to cleanse arm before collecting any specimen for volatile testing.

2. Avoid exposure of specimen to atmosphere.

3. Within 2 hours of collection, centrifuge the specimen.

4. For serum gel tubes,  do not aliquot. Send centrifuged specimen in original tube.

5. For red-top tubes, aliquot serum into a plastic vial.


Specimen Minimum Volume

Serum: 0.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 14 days
  Frozen  28 days
  Ambient  24 hours

Reference Values

Methanol:

Not detected (Positive results are quantitated.)

Toxic concentration: ≥10 mg/dL

 

Ethanol:

Not detected (Positive results are quantitated.)

Toxic concentration: ≥400 mg/dL

 

Isopropanol:

Not detected (Positive results are quantitated.)

Toxic concentration: ≥10 mg/dL

 

Acetone:

Not detected (Positive results are quantitated.)

Toxic concentration: ≥10 mg/dL

Day(s) Performed

Monday through Saturday

CPT Code Information

G0480

80320 (if appropriate for select payers)

Clinical Toxicology CPT Code Client Guidance

Report Available

1 to 2 days

Reject Due To

Gross hemolysis OK
Gross lipemia OK
Gross icterus OK
 

Specimen Retention Time

2 weeks