Sign in →

Test Code VALP Valproic Acid, Total, Plasma

Additional Codes

LAB24

Performing Laboratory

Central Peninsula Hospital

Important Notes

Valproic Acid, Total, Plasma test is for medical use only.

Collection Instructions

Draw trough levels immediately prior to next dose.

Collection Method

Venipuncture

Container

Lt Green / Plasma Separator / Lithium Heparin

Specimen Type

Preferred

  • Plasma- Light Green Top

 

Acceptable

  • Serum- Gold Top SST or Red Top

Minimum Volume

0.5ml

Specimen Stability

Ambient 2 days
Refrigerated (4°C to 8°C) 7 days
Frozen(-40°C to 0°C) 3 months

Reference or Target Ranges

50 – 100 μg/mL

 

Peak: 3 - 4 hours post oral dose

Trough: Just prior to next dose

Reportable Units

μg/ml

Critical Value

trough ≥ 150 µg/ml

Reasons for Rejection

Hemolysis Threshold 500 mg/dL
Icterus Threshold 30 mg/dL
Lipemia Threshold 500 mg/dL
Other QNS, contaminated, improperly labeled, improper specimen type

Performance Information

Days and Time Performed 24/7
Expected Turn Around Time Stat: 1 hour
Routine: 4 hours
Stat Availability Yes
Performing Bench CPH Chemistry
Methodology/Method Description Roche Cobas 

CPT Codes

80164

Last review/edit date

2025