Sign in →

Test Code POC DAU Point of Care Drugs of Abuse 14 with Adulterants

Additional Codes

POC185

Performing Laboratory

Central Peninsula Hospital

Collection Method

Urine specimen should be collected in an iScreen test cup

Container

iscreen test cup

Specimen Type

Sterile Urine Cup

 

Minimum Volume

A minimum amount of urine must be collected as indicated by the fill line on the test cup

Specimen Stability

Perform as soon as possible

Reference or Target Ranges

Negative

 

Drug Result Interpretation Chart

Adulterant Guide

 

Drug Cut-off Level
Amphetamine (AMP) 500 ng/mL
Secobarbital (BAR) 300 ng/mL
Buprenorphine (BUP) 10 ng/mL
Oxazepam (BZO) 300 ng/mL
Cocaine (COC) 150 ng/mL
Fentanyl (FTY) 1 ng/mL
Methylenedioxymethamphetamine (MDMA) 500 ng/mL
Methamphetamine 500 ng/mL
Morphine (MOP/OP) 300 ng/mL
Methadone (MTD) 300 ng/mL
Oxycodone (OXY) 100 ng/mL
Phencyclidine (PCP) 25 ng/mL
Nortriptyline (TCA) 1000 ng/mL
Marijuana (THC) 50 ng/mL

Reportable Units

ng/mL

Critical Value

Not defined for this assay

Reasons for Rejection

QNS, contaminated, improperly labeled, improper specimen type

Performance Information

Days and Time Performed 24/7
Performing Bench CPH Point of Care
Methodology/Method Description iScreen DX Drug Screen
Tox Cup

CPT Codes

80305

Last review/edit date

2025