Test Code GLU PP Glucose, 2hr Post Prandial
Additional Codes
LAB13108
Collection Method
Venous collection
Minimum Volume
0.5ml
Container
Lt Green / Plasma Separator / Lithium Heparin
Acceptable Specimens
Serum:
- Gold Top SST
- Red Top
Plasma:
- Light Geen Top (lithium heparin)
Reference or Target Ranges
65-139 mg/dL
According to the American Diabetes Association, a glucose value between 140 and 199 mg/dL is consistent with impaired glucose tolerance. A glucose value equal to or greater than 200 mg/dL is consistent with a diagnosis of Diabetes Mellitus if confirmed with another test.
Reportable Units
mg/dl
Critical Value
<50 mg/dL or >500 mg/dL
Reasons for Rejection
Hemolysis Threshold | 1000 mg/dL |
Icterus Threshold | 60 mg/dL |
Lipemia Threshold | 1000 mg/dL |
Other | QNS, contaminated, improperly Labeled, improper specimen type |
Specimen Stability
Ambient | 8 hours |
Refrigerated (4°C to 8°C) | 3 days |
Frozen(-40°C to 0°C) | 3 days |
Performance Information
Days and Time Performed | 24/7 |
Expected Turn Around Time |
Stat: 1 hour Routine: 4 hours |
Stat Availability | Yes |
Performing Bench | CPH Chemistry |
Methodology/Method Description | Roche Cobas |
CPT Codes
82950
Last review/edit date
2025