Test Code CRP C-Reactive Protein (CRP), Plasma or Serum
Additional Codes
LAB149
Collection Method
Venous collection
Minimum Volume
0.5ml
Container
Lt Green / Plasma Separator / Lithium Heparin
Acceptable Specimens
Serum:
- Gold Top SST
- Red Top
Plasma:
- Light Geen Top (lithium heparin)
Reference or Target Ranges
<0.8 mg/dl
Reportable Units
mg/dL
Critical Value
Not defined for this assay
Reasons for Rejection
Hemolysis Threshold | 1000 mg/dL |
Icterus Threshold | 60 mg/dL |
Lipemia Threshold | 1000 mg/dL |
Other | QNS, contaminated, improperly Labeled, improper specimen type |
Specimen Stability
Ambient | 11 days |
Refrigerated (4°C to 8°C) | 2 months |
Frozen(-40°C to 0°C) | 3 years |
Performance Information
Days and Time Performed | 24/7 |
Expected Turn Around Time | Stat: 1 hour Routine: 4 hours |
Stat Availability | Yes |
Performing Bench | CPH Chemistry |
Methodology/Method Description | Roche Cobas |
CPT Codes
86140
Last review/edit date