Sign in →

Test Code BF AMY Amylase, Body Fluid

Additional Codes

LAB178

Collection Instructions

Source must be noted

Collection Method

Sterile body fluid collection

Container

Sterile Container, Syringe or Red top tube

Specimen Type

Preferred- Sterile Container, Syringe or Red top tube

Acceptable- Sterile Container, Syringe or Red top tube

Minimum Volume

0.5ml

Specimen Stability

Ambient 7 days
Refrigerated (4°C to 8°C) 1 month
Frozen(-40°C to 0°C) N/A

Reference or Target Ranges

No reference values available for body fluid sources

Reportable Units

U/L

Reasons for Rejection

Hemolysis Threshold 200 mg/dL
Icterus Threshold 60 mg/dL
Lipemia Threshold 1500 mg/dL
Other QNS, contaminated, improperly Labeled, improper specimen type

Performance Information

Days and Time Performed 24/7
Expected Turn Around Time Stat: 1 hour
Routine: 4 hours
Stat Availabilty Yes
Performing Bench CPH Chemistry
Methodology/Method Description Roche Cobas 

CPT Codes

82150

Last review/edit date

2025