Test Code BF ALB Albumin, Body Fluid
Additional Codes
LAB177
Collection Method
Sterile body fluid collection
Minimum Volume
0.5ml
Container
Sterile Container, Syringe or Red top tube
Acceptable Specimens
Body fluids
Reference or Target Ranges
0.2-6.0 g/dL
Reportable Units
g/dL
Critical Value
Not defined for this assay
Reasons for Rejection
Hemolysis Threshold | 1000 mg/dL |
Icterus Threshold | 60 mg/dL |
Lipemia Threshold | 550 mg/dL |
Other | QNS, contaminated, improperly Labeled, improper specimen type |
Specimen Stability
Ambient | 2.5 months |
Refrigerated (4°C to 8°C) | 5 months |
Frozen(-40°C to 0°C) | 4 months |
Performance Information
Days and Time Performed | 24/7 |
Expected Turn Around Time | Stat: 1 hour Routine: 4 hours |
Stat Availabilty | Yes |
Performing Bench | CPH Chemistry |
Methodology/Method Description | Roche Cobas |
CPT Codes
82042
Last review/edit date
Feb 2024