Sign in →

Test Code AL Aluminum, Serum

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Patient Preparation: High concentrations of gadolinium and iodine are known to interfere with most metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.

Supplies:

-Metal Free B-D Tube (No Additive), 6 mL (T184)

-Metal Free Specimen Vial (T173)

Container/Tube: 6-mL Plain, royal blue-top Vacutainer plastic trace element blood collection tube

Submission Container/Tube: 7-mL Metal-free, screw-capped, polypropylene vial

Specimen Volume: 1.2 mL

Collection Instructions: See Metals Analysis Specimen Collection and Transport for complete instructions.


Specimen Minimum Volume

0.3 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 7 days METAL FREE
  Ambient  7 days METAL FREE
  Frozen  7 days METAL FREE

Reference Values

<7 ng/mL

<60 ng/mL (dialysis patients)

 

For International System of Units (SI) conversion for Reference Values, see www.mayocliniclabs.com/order-tests/si-unit-conversion.html

Day(s) Performed

Tuesday

CPT Code Information

82108

Report Available

2 to 8 days

Specimen Retention Time

14 days

Reject Due To

Gross hemolysis OK
Gross lipemia OK