Sign in →

Test Code POC COOX Point of Care Oximeter

AVOXimeter User Manual

AVOXimeter User Manual

Collection Method

  • Whole blood collected in a heparinized syringe
  • EDTA preserved specimens may also be used

Minimum Volume

50ul

Container

  • Whole blood collected in a heparinized syringe
  • EDTA preserved specimens may also be used

Reference or Target Ranges

  Age Male Female
THb 12 to 17 years old 13 - 16 g/dL 13 - 16 g/dL
THb 18 years and older 14 - 17.7 g/dL 12 - 15.7 g/dL
O2Hb Arterial All ages 95 - 99 % 95 - 99 %
O2Hb Venous All ages 60 - 85 % 60 - 85 %

 

Expected Results:

Source %O2Hb
PA 73-85%
PCW 90-100%
RV 64-84%
RA 72-86%
ICV 76-88%
SVC 66-84%

 

Reportable Units

Varies

Critical Value

THb -  <7 or >18 g/dL

 

Reasons for Rejection

QNS, contaminated, improperly Labeled, improper specimen type

Specimen Stability

  • Samples should be analyzed immediately.
  • Samples collected with heparin or EDTA preservative can be kept in a capped plastic syringes for up to 30 minutes at room temperature.

Acceptable Specimens

  • Whole blood collected in a heparinized syringe
  • EDTA preserved specimens may also be used

Performance Information

Days and Time Performed 24/7
Performing Bench CPH Point of Care
Methodology/Method Description AVOXimeter

CPT Codes

82810

Last review/edit date

2025